An automatic digital polarimeter

A polarimeter [1] is a scientific instrument used to measure the angle of rotation caused by passing polarized light through an optically active substance.[2]

Some chemical substances are optically active, and polarized (unidirectional) light will rotate either to the left (counter-clockwise) or right (clockwise) when passed through these substances. The amount by which the light is rotated is known as the angle of rotation.

Contents

History [edit]

Polarization by reflection was discovered in 1808 by Étienne-Louis Malus (1775–1812).[2]

Construction [edit]

The polarimeter is made up of two Nicol prisms (the polarizer and analyzer). The polarizer is fixed and the analyzer can be rotated. The prisms may be compared to as slits S1 and S2. The light waves may be considered to correspond to waves in the string. The polarizer S1 allows only those light waves which move in a single plane. This causes the light to become plane polarized. When the analyzer is also placed in a similar position it allows the light waves coming from the polarizer to pass through it. When it is rotated through the right angle no waves can pass through the right angle and the field appears to be dark. If now a glass tube containing an optically active solution is placed between the polarizer and analyzer the light now rotates through the plane of polarization through a certain angle, the analyzer will have to be rotated in same angle.

Operation [edit]

Polarimeters measure this by passing monochromatic light through the first of two polarising plates, creating a polarized beam. This first plate is known as the polarizer.[3] This beam is then rotated as it passes through the sample. The sample is usually prepared as a tube where the optically active substance is dissolved in an optically inactive chemical such as distilled water, ethanol, methanol. Some polarimeters can be fitted with tubes that allow for sample to flow through continuously.

After passing through the sample, a second polarizer, known as the analyzer, rotates either via manual rotation or automatic detection of the angle. When the analyzer is rotated to the proper angle, the maximum amount of light will pass through and shine onto a detector.

Types of polarimeter [edit]

Laurent's half-shade polarimeter [edit]

Biquartz polarimeter [edit]

Lippich polarimeter [edit]

Manual [edit]

The earliest polarimeters, which date back to the 1830s, required the user to physically rotate one polarizing element (the analyzer) whilst viewing through another static element (the detector). The detector was positioned at the opposite end of a tube containing the optically active sample, and the user used his eye to judge the "alignment" when least light was observed. The angle of rotation was then read from a simple protractor fixed to the moving polariser to within a degree or so.

Although most manual polarimeters produced today still adopt this basic principle, the many developments applied to the original opto-mechnical design over the years have significantly improved measurement performance. The introduction of a half-wave plate increased "distinction sensitivity", whilst a precision glass scale with vernier drum facilitated the final reading to within ca. ±0.05º. Additionally, most modern day manual polarimeters also incorporate a long-life yellow LED in place of the more traditional and costly sodium arc lamp.

Semi-automatic [edit]

Today there are also semi-automatic polarimeters, which require visual detection but use push-buttons to rotate the analyzer and offer digital displays.

Fully automatic [edit]

The most modern polarimeters are fully automatic, and simply require the user to press a button and wait for a digital readout. Fast automatic digital polarimeters reduce measuring time to just one second, regardless of the rotation angle of the sample. In addition, they permit continuous measurement, for example for kinetic investigations or in HPLC. Special techniques like a temperature controlled sample tube reduce measuring faults and ease operation. Results can directly be transferred to computers or networks for automatic processing.[4]

Sources of error [edit]

The angle of rotation of an optically active substance can be affected by:

  • Concentration of the sample
  • Wavelength of light passing through the sample (generally, angle of rotation and wavelength tend to be inversely proportional)
  • Temperature of the sample (generally the two are directly proportional)
  • Length of the sample cell (input by the user into most automatic polarimeters to ensure better accuracy)

Most modern polarimeters have methods of compensating for or controlling these errors.

Calibration [edit]

Polarimeters can be calibrated – or at least verified – by measuring a quartz plate, which is constructed to always read at a certain angle of rotation (usually +34°, but +17° and +8.5° are also popular depending on the sample). Quartz plates are preferred by many users because solid samples are much less affected by variations in temperature, and do not need to be mixed on-demand like sucrose solutions.[5]

Applications [edit]

Because many optically active chemicals such as sucrose, are stereoisomers, a polarimeter can be used to identify which isomer is present in a sample – if it rotates polarized light to the left, it is a levo-isomer, and to the right, a dextro-isomer. It can also be used to measure the ratio of enantiomers in solutions.

Chemical industry [edit]

Many chemicals exhibit a specific rotation as a unique property (like refractive index in many cases) which can be used to distinguish it. Polarimeters can identify unknown samples based on this if other variables such as concentration and length of sample cell length are controlled or at least known. This is used in the chemical industry.

By the same token, if the specific rotation of a sample is already known, then the concentration and/or purity of a solution containing it can be calculated.

Most automatic polarimeters make this calculation automatically, given input on variables from the user.

Food, beverage and pharmaceutical industries [edit]

Concentration and purity measurements are especially important to determine product or ingredient quality in the food & beverage and pharmaceutical industries. Samples that display specific rotations that can be calculated for purity with a polarimeter include:

Polarimeters are used in the sugar industry for determining quality of both juice from sugar cane and the refined sucrose. Often, the sugar refineries use a modified polarimeter with a flow cell called a saccharimeter. These instruments use the International Sugar Scale (as defined by the International Commission for Uniform Methods of Sugar Analysis (ICUMSA).

References [edit]

See also [edit]


Original courtesy of Wikipedia: http://en.wikipedia.org/wiki/Polarimeter — Please support Wikipedia.
A portion of the proceeds from advertising on Digplanet goes to supporting Wikipedia.
379 videos foundNext > 

The Principle of Polarimetry

This lesson demonstrates how polarizers work to detect chiral molecules in solution.

Polarimeter - Yale CBIC

Training video on polarimetry by Eric Paulson. DISCLAIMER: The information provided in the video is not a substitute for direct hands-on training. Do not att...

POLARIMETER MOD.PL 1 ( OPTICS LAB) IRAQ

IRAQ / Qadissiya University / College of Education / Physics Department / Optics Lab .. polarimeter can be used in the chemical industry, hospitals, colleges...

Polarimeter, Polarimetry, Laurent Half-Shade Polarimeter

POLARIMETER The Polarimeter consists of Monochromatic source of light usually a sodium lamp. This light is allowed to pass through a narrow slit and then to ...

20.6.6 Outline the use of a polarimeter in distinguishing between optical isomers IB Chemistry

Plane polarized light will rotate in different directions when passing through different stereo isomers (from a pair of enantiomers). A polarimter measures t...

Analyse mit dem Polarimeter | Chemie | Analytische Chemie

Komplettes Chemie-Video unter http://www.sofatutor.com/v/2ew/GP In diesem Video geht es um die Analyse mit dem Polarimeter. Dazu werden die Grundlagen der Me...

Polarimeter Measurement with an Autopol from Rudolph Research Autopol

Polarimeter Technical Video: Taking a measurement with Rudolph Research's Autopol V Plus, Digital, Bench Top Polarimeter. The proper procedures for loading a...

Polarimeter Experiment

Made by Patrick Jung and Megan Para. Featuring James Liu in the background.

Home made polarimeter, #2

The video shows a home made polarimeter with 4 "chambers" made from 4 oz fruit cans glued to a pane of glass. The glass pane is resting on the screen of a la...

Automatic Polarimeter SEPA-500 [English]

The features and usage examples of the newly redesigned SEPA-500 automatic polarimeter are introduced in this video.

379 videos foundNext > 

6 news items

PR Web (press release)

PR Web (press release)
Tue, 14 May 2013 00:14:11 -0700

ATAGO released its state of the art SAC-i Automatic Polarimeter/Saccharimeter in April of 2013. It has already received rave reviews from sugar producers around the world. The SAC-i has been a big hit especially in Brazil, the sugar production capital ...
 
Labmate Online
Mon, 22 Apr 2013 16:10:14 -0700

Krüss are very proud of their Polarimeter P8000 because it's the fastest polarimeter worldwide. Krüss developed a high-speed measurement that reduces measuring time of samples to only 1 second irrespective of the angle of the sample. Compared to other ...
 
Science Daily (press release)
Wed, 24 Apr 2013 05:18:58 -0700

(1) An instrument used for observations is the IRSF (InfraRed Survey Facility), a 1.4 m telescope with a built-in SIRPOL imaging polarimeter at an elevation of about 1761 meters at the South African Astronomical Observatory in Sutherland, South Africa.

Media Inaf

Media Inaf
Tue, 07 May 2013 03:42:45 -0700

Il Polarimeter Add-On for the LRS Optics, uno strumento in grado di compiere misure di polarizzazione in ottico. "In astrofisica studi di polarizzazione sono cruciali in diversi settori" spiega Stefano Covino che insieme a Emilio Molinari ha guidato il ...

Dire (Abbonamento)

Dire (Abbonamento)
Fri, 10 May 2013 05:47:35 -0700

A realizzare il Polarimeter Add-On for the LRS Optics, questo il nome esteso, è stato un gruppo di ricercatori dell'Inaf. Il suo compito è estrarre dalla radiazione elettromagnetica proveniente dai corpi celesti non solo le informazioni legate alla sua ...

OggiScienza

OggiScienza
Thu, 09 May 2013 05:21:09 -0700

Su un'isola dell'arcipelago delle Canarie è da poco pienamente operativo un nuovo strumento: il Polarimeter Add-On for the LRS Optics, altrimenti detto PAOLO. PAOLO è un polarizzatore, cioè un filtro che seleziona la radiazione elettromagnetica a ...
Loading

Oops, we seem to be having trouble contacting Twitter

Talk About Polarimeter

You can talk about Polarimeter with people all over the world in our discussions.

Support Wikipedia

A portion of the proceeds from advertising on Digplanet goes to supporting Wikipedia. Please add your support for Wikipedia!